6. BASE 120 / BASE 600 Firmware Update

6.1. Preparation

Connect your BASE module to your PC. Start the software Vci3floadGui or Vci4floadGui via Windows Explorer. You will find the software in C:\Program Files\IXXAT\3.5 or C:\Program Files\HMS\IXXAT VCI 4.0 after the installation of the QmixElements or Nemesys User Interface software (see figure below).

_images/image98.png

The software will show you the detected USB-to-CAN interface. If the software does not detect any USB-to-CAN interface then please try to connect your BASE module to another USB interface of your PC.

You only need to do an update, if you have a V2 device, that means, if you have an USB-to-CAN V2 embedded or an USB-to-CAN V2 compact. If you do not have a V2 device, then you are finished now.

6.2. Update procedure

An USB-to-CAN V2 device should appear in the Device field of the Flashloader window.

_images/image99.png

Click the >>` button to select the firmware (e.g. USB-to-CAN_V2_FW_0_1_6_3_R380.hex). You will find the latest firmware here:

https://CETONI.de/software-downloads/

Attention

Do not interrupt the process while the flash procedure is running!

Now click the Flash button to start the update process . Wait for the process to complete and then click Verify. If the verification succeeds with an OK message (see Figure below) the update process is complete and your device has been updated successfully.

_images/image100.png